Portable Ventilator
MCV 200
ALLIED Healthcare Products INC
 
 
 
Can be operated solely on compressed gas, DC power, or AC power, offering the greatest flexibility for mass casualty responder
 
Simple control interface allows for just-in-time training and use by a wide range of responders
Assist-Control, external PEEP, and independent ventilation controls accommodate enhanced treatment options
Low maintenance and weatherproof design ensure low cost of ownership in stockpile or in everyday use
 
The MCV200 provides an additional level of preparedness when planning for an unknown
 
medical surge. In addition to its internal battery and compressors that provide the flexibility
 
of operationon AC or extended operation on DC power, the MCV200 also incorporates a completely pneumatic ventilation mode ensuring operation even when electricity is unavailable and all DC power sources are drained. The MCV200 is capable of providing ventilation using ambient air, high-pressure oxygen, or an oxygen/air mix. Like the MCV100, the MCV200
 
features intuitive ventilation controls and an array of adjustable alarms to enhance patient safety. The unique power source flexibility of the MCV200 combined with a rugged, weather-resistant and low-maintenance design ensures the greatest level of preparedness for mass casualty planners
 
 
Specifications - MCV200
 
 
Setting Parameters
 
 
ventilation Mode: Assist-Control
 
Assist-COntrol: Triggers at less than -2 cm H20
 
Flow Rate: 12-36 LPM
 
Breathing Rate: Adjustable. 0, 8-20 breaths per minute
 
TIdal volume: Adjustable. 200-1200 m1
 
Inspiratory TIme: 1 or 2 seconds
 
PEEP: External. 0-20 cm H20 (with PEEP adapter)
 
Fi02: Selectable. 21 %, 65%, 100%
 
Manometer Readout: Digital. 0-60 cm H20
 
Airway Pressure Limit: Adjustable. 5-60 cm H20
 
 
Power Characteristics
 
 
Power Source: Compressed O2, 110V-240V AC, internal battery
 
external12V DC battery
 
O2 Supply Pressure: 40-87 psi
 
 
Operating TIme on Battery
 
 
Based on average adult settings of 10 BPM, 600 m1 tidal volume
 
2 second inspiratory time
 
Air Mode (using internal compressors): Approx. 7 hours
 
100% O2 or 65% O2 Mode: Approx. 21 hours
 
Internal Battery: 12 Volt 5 Ah Rechargeable. Sealed lead acid type
 
Battery Recharge TIme: 5 hours (less than 24 hours when in use)
 
Battery Shelf Life: 3 years
 
 
Alarms  
 
High Airway Pressure: Adjustable. 15-60 cm H20
 
Low Airway PressurelDisconnect: Adjustable. 0-30 cm H20
 
Low Source Gas: Fixed. Activates at about 40 psi (275 kPa) Temperature
 
Low Battery: Activates when approx. 1Y.2 hours of run time remain
 
Alarm Volume: Approx. 60 dB at 1 meter
 
Alarm Silence: Yes. 110 Seconds
 
 
Physical Characteristics
 
 
Dimensions: 3.5" x 10.3" x 14.5" (88.9 mm x 261.1 mm x 368.3 mm)
 
Weight: 17.2Ibs. (7.8 kg)
 
 
Temperature Ranges
 
 
Operating: -15 to 122°F (-9 to 50°C)
 
Storage: -40 to 140°F (-40 to 60°C)
 
Enclosure: ABS plastic with seals to prevent water ingress
 
Shock Resistance: 30" drop
 
Vibration Resistance: 100 G
 
Water Resistance: Spill and rain resistant